  • LISN 50uH - Line Impedance Stabilization Network

Line Impedance Stabilization Network for conducted EMC measurements

V-Type LISN 50 / 250uH 16A


For development and pre-compliance measurements

For pre-compliance conducted EMI testing an Artificial Main or Line Impedance Stabilization Network is required. The LISN 50/16 AC is a measurement platform that is compliant with CISPR 16-1-2, VDE 0876 and FCC part 15.

A LISN is used to safely separate the 9kHz - 110MHz band from the line voltage and provides a standardized impedance over which is measured.
It can be used as a V-, Δ- and T-network, which makes it ideal for solving EMI problems.

Find a nearby Distributor.

LISN 50uH - Line Impedance Stabilization Network  | ELab Tools

Conducted EMI

Using the built-in DM, CM and single line measurement, the LISN50/16 eases problem analysis to find the noise source in a system.

Filtered Mains

The 250μH input (pre-)filter eliminates interference from the mains outlet.

Low/High pass

The selectable low- and high-pass filters eliminate measurement errors due to clipping of the measurement receiver. The filters are combined with a built-in attenuator/limiter that protects the spectrum analyzer from harmfull input transients during power on of the equipment under test.

This measurement instrument has been designed by electronic engineers for electronic engineers who deal with devices that are connected to the Mains like inverters, machines and power supplies.

Use

The circuit used, conforms to the specifications of EN 55016-1-2, VDE 0876 and FCC part 15 with V-, Δ- and T-network measurements.

The simulated impedance is (50μH / 250μH + 5 Ω ) || 50 Ω in the frequency range 9 kHz to 110 MHz. Filters are provided to reduce the bandwidth.

It is designed to allow for peaks up to 30A without loss of linearity.

Note: Due to the large Y capacitors, an isolation transformer or floating power source is recommended/required.

The LISN 50/16 features in brief

  • 50μH / 250μH + 5Ω network 16A
  • L or N
  • Common mode (CM)
  • Differential mode (DM)
  • Artificial hand grounding
  • Low-pass filter 35MHz
  • High-pass filter 9kHz
  • Built-in limiter
  • Measurement range 9kHz - 110MHz
  • Attn 10dB

Suitable for measurement like

  • CISPR 16-1-2
  • VDE 0876
  • FCC part 15
  • EN61000-6-3/4
  • EN 55016-1-2